What if the KIC furnace temperature tester cannot read data from the furnace?

How to reset the instrument after automatic overheating protection?


Exit KIC software

1. Find KIC Profiler 2000 \ Log \ kic2000usersettings.kiccfg and open it in Notepad mode

2. Change ViewHostDialogue=0 to=1 in the file, then save and exit.

Start the KIC2000 software again and you will see the KICHost interface.

Pay attention to the following operations, preferably in English version

Remember that the KIC furnace temperature tester uses a 9V battery


1. Make sure the SlinKIC2000 software is close.
2. Under the root directory on your hard drive, under the folder the KIC Profiler 2000 you will find a folder called "Log".(Using Windows Explorer, open the X:\KIC Profiler 2000\Log\kic2000usersettings.kiccfg file)
3. Under the Log folder there is a file called "KIC2000UserSettings.KICCFG"
4. Open this file using Note or WordPad text editor.
5. With file opened under the text editor, find the line:  ( " ViewHostDialog=0", Set this line to =1, Save changes to the file. )
6. Restart the SlimKIC2000 software and you motice that a second application will open, the KIC Host. ( On the Windows Taskbar, click on the "KIC Host" tab)
7. Direct connection to SlimKIC2000, restart SlimKIC2000 Unit.
8. Select "Configure SlimKIC" .
9. Another Window will open. Select "Write SlimKIC Setuop" . You should see dialog appear in that window which says " Write SlimKIC Successful".